I have a question about the scale limit setting using the ocx component FLIR SDK Camera Control V 1.0 (Thermovision SDK 2.5 SR-1)
I would change the scale limits in camera acquisition
The panel Settings (LVCam panel->Tab Dev->Settings... button) works fine.
But when I use the Metod
SetCameraProprierty(8,value) and
SetCameraProprierty(9,value),
to programmatically set them, both the limit values are -273.2.
What are the right propertyID's to use?
Solution:
SDK 2.5x Camera Control Update
Camera Control Update for SDK 2.5 Version: 2.90.10
Download and unzip. Before installation, open readme file!
Last Updated: March 18, 2008
Camera Control Update for SDK 2.5 Version: 2.90.10
Download and unzip. Before installation, open readme file!
Last Updated: March 18, 2008